Autoclave

Autoclave

Equipment number : 4864

Autoclave

Manufacturer: Getinge

Model: HS 6613 AM-2

Year: 2004

Volume: 585 liters

Basket available

Overall dimensions (l x d x h): 1800 x 1200 x 2000 mm

Weight: 1200 kg

Working dimensions (l x d x h): 672 x 672 x 1300 mm

Fluid: steam

Working pressure: 3 bar

Test pressure: 5 bar

Working temperature: + 20 / + 148 °C

Double-jacket: 6 x 18 liters / working pressure 3,5 bar / test pressure 7,5 bar / working temperature + 20 / + 148 °C

EC certificate says:

Category III equipment

Design pressure: – 1 / 3,5 bar

Design temperature: + 20 / + 148 °C

The qualification file indicates a conform vacuum test

Control unit: PACS 3000 (Programmable Autoclave control System)

Program specification B3113

Prom version: H32086 (B3113. 2,1)

System version: V 3,09 Fr.

Technical documentation includes: layout plan / installation data / user manual / electrical diagrams / maintenance manual / user manual for the linear recorder / passwords / pressure devices file / test cycle records / adjustment certificates